GORE® STA-PURE® SU PRO Transfer Pump Tubing
Reduce risk in demanding single-use bioprocessing with radiation-compatible, flex-durable pump tubing engineered for longer run times and high pressures.
Confidence for Demanding Single-Use Bioprocessing
As continuous and intensified bioprocessing increases run times and operating demands, pump tubing can become a vulnerable component in the fluid path. GORE STA-PURE SU PRO Transfer Pump Tubing is designed to help reduce tubing-related risks while supporting extended fluid transfers in upstream and downstream single-use applications.
Its unique composite construction combines platinum-cured silicone with expanded polyethylene (ePE) reinforcement for added strength and resilience. Drawing on our materials science and pump tubing expertise, GORE STA-PURE SU PRO tubing is engineered for demanding pumping conditions while supporting radiation-compatible single-use workflows.
Featured Resources

Why We Developed GORE STA-PURE SU PRO Pump Tubing
Hear from a Gore engineer about the changing demands of single-use bioprocessing and the engineering behind GORE STA-PURE SU PRO Transfer Pump Tubing.
Technical Specifications
Unlike conventional tubing, GORE STA-PURE SU PRO Transfer Pump Tubing is a composite of platinum cured silicone, reinforced with expanded polyethylene (ePE) for added strength and resilience (patent pending).
Product Information
| Materials and Construction | Composite of thermoplastic material (Expanded polyethylene (ePE) membrane) and elastomeric materials (platinum-cured silicone) |
|---|---|
| Fluid contact surface | Platinum-cured silicone |
| Performance Testing |
GORE STA-PURE SU PRO tubing was tested for durability in peristaltic pumps over a wide range of discharge pressures and working fluids (see table).
| Sterilization |
|
|---|---|
| Biocompatibility |
|
| Manufacturing Environment |
|
| Quality Standards |
|
| Inspection |
|
| Storage/Shelf Life |
|
Testing conditions
- Continuous pumping of room temperature deionized water in Watson-Marlow 500 Series peristaltic pumps, with pump head speed of 220 rpm (two rollers)
- All GORE STA-PURE SU PRO samples were exposed to gamma radiation in the range of 45 – 56 kGy
- Tube life was measured as time until failure. Tube failure was defined as rupture, leakage, sustained flow reduction of 25 percent or more, or inability to sustain operating pressure.
NOTES (testing at elevated pressure)
- Sample size of six (6) for each tube type
- The data is a representative sampling, may not be indicative of future results, and does not constitute creation of a specification
- Nominal tube size: 6.4 mm (inner diameter) x 2.4 mm (wall thickness)
- Minimum observed tube life for an individual GORE STA-PURE SU PRO sample at 4 bar (60 psig): 415 hours
- Minimum observed tube life for an individual TPE sample at 4 bar (60 psig): 1.6 hours
- All TPE samples were not irradiated
- Average tube life for STA-PURE SU PRO Tubing in Figure 1 is based on 3 of 6 samples reaching end-of-test due to tube failure. The remaining 3 samples performed without failure until the test was terminated at 504 hours (21 days).
GORE STA-PURE SU PRO products are available in standard configurations, sized for easy retrofit with industry-standard equipment. See a complete list of all our standard dimensions. To request a Validation Guide, or to connect with a trusted channel partner to discuss your specific needs, contact us.
Part Number Description
Example | Z | B | 14 | RC |
|---|---|---|---|---|
Attribute | Inner diameter (ID) | Wall thickness (WT) | Length (L) | Radiation Compatible |
Nominal Range | 1.6 – 25.4 mm (0.063 – 1.0 in) | 1.6 – 4.8 mm (0.063 – 0.189 in) | 208.0 – 660.4 mm (8.2 – 26.0 in) |
|
Example: ZB14RC is STA-PURE SU PRO Tubing with a nominal ID of 3.2 mm (0.13 in.), WT of 2.4 mm (0.094 in.), and L of 355.6 mm (14 in.)
| Nominal Tube Size | Dimensional Specifications | |||||
|---|---|---|---|---|---|---|---|
Gore Part Number | Inner Diameter (mm) | Wall Thickness (mm) | Inner Diameter (mm) | Wall Thickness (mm) | Length (mm) | ||
YA12RC | 1.6 | 1.6 | 1.56 ± 0.20 | 1.70 ± 0.20 | 304.8 ± 3.0 | ||
YA14RC | 1.6 | 1.6 | 1.56 ± 0.20 | 1.70 ± 0.20 | 355.6 ± 3.0 | ||
ZA12RC | 3.2 | 1.6 | 3.04 ± 0.20 | 1.70 ± 0.20 | 304.8 ± 3.0 | ||
ZA14RC | 3.2 | 1.6 | 3.04 ± 0.20 | 1.70 ± 0.20 | 355.6 ± 3.0 | ||
AA12RC | 4.8 | 1.6 | 4.58 ± 0.25 | 1.70 ± 0.20 | 304.8 ± 3.0 | ||
AA14RC | 4.8 | 1.6 | 4.58 ± 0.25 | 1.70 ± 0.20 | 355.6 ± 3.0 | ||
BA12RC | 6.4 | 1.6 | 6.07 ± 0.25 | 1.70 ± 0.20 | 304.8 ± 3.0 | ||
BA14RC | 6.4 | 1.6 | 6.07 ± 0.25 | 1.70 ± 0.20 | 355.6 ± 3.0 | ||
BA16RC | 6.4 | 1.6 | 6.07 ± 0.25 | 1.70 ± 0.20 | 406.4 ± 3.0 | ||
BA17RC | 6.4 | 1.6 | 6.07 ± 0.25 | 1.70 ± 0.20 | 431.8 ± 3.0 | ||
BA24RC | 6.4 | 1.6 | 6.07 ± 0.25 | 1.70 ± 0.20 | 609.6 ± 3.0 | ||
CA12RC | 8.0 | 1.6 | 7.59 ± 0.25 | 1.70 ± 0.20 | 304.8 ± 3.0 | ||
CA14RC | 8.0 | 1.6 | 7.59 ± 0.25 | 1.70 ± 0.20 | 355.6 ± 3.0 | ||
CA16RC | 8.0 | 1.6 | 7.59 ± 0.25 | 1.70 ± 0.20 | 406.4 ± 3.0 | ||
YB8RYC* | 1.6 | 2.4 | 1.52 ± 0.20 | 2.50 ± 0.20 | 208.0 ± 3.0 | ||
YB14RC | 1.6 | 2.4 | 1.52 ± 0.20 | 2.50 ± 0.20 | 355.6 ± 3.0 | ||
ZB8RYC* | 3.2 | 2.4 | 3.03 ± 0.25 | 2.50 ± 0.20 | 210.1 ± 3.0 | ||
ZB14RC | 3.2 | 2.4 | 3.03 ± 0.25 | 2.50 ± 0.20 | 355.6 ± 3.0 | ||
AB8RYC* | 4.8 | 2.4 | 4.62 ± 0.25 | 2.50 ± 0.20 | 210.1 ± 3.0 | ||
AB14RC | 4.8 | 2.4 | 4.62 ± 0.25 | 2.50 ± 0.20 | 355.6 ± 3.0 | ||
AB16RC | 4.8 | 2.4 | 4.62 ± 0.25 | 2.50 ± 0.20 | 406.4 ± 3.0 | ||
AB26RC | 4.8 | 2.4 | 4.62 ± 0.25 | 2.50 ± 0.20 | 660.4 ± 3.0 | ||
BB8RYC* | 6.4 | 2.4 | 6.19 ± 0.25 | 2.50 ± 0.20 | 214.9 ± 3.0 | ||
BB14RC | 6.4 | 2.4 | 6.19 ± 0.25 | 2.50 ± 0.20 | 355.6 ± 3.0 | ||
BB16RC | 6.4 | 2.4 | 6.19 ± 0.25 | 2.50 ± 0.20 | 406.4 ± 3.0 | ||
BB24RC | 6.4 | 2.4 | 6.19 ± 0.25 | 2.50 ± 0.20 | 609.6 ± 3.0 | ||
BB26RC | 6.4 | 2.4 | 6.19 ± 0.25 | 2.50 ± 0.20 | 660.4 ± 3.0 | ||
CB8RYC* | 8.0 | 2.4 | 7.75 ± 0.25 | 2.50 ± 0.20 | 214.9 ± 3.0 | ||
CB14RC | 8.0 | 2.4 | 7.75 ± 0.25 | 2.50 ± 0.20 | 355.6 ± 3.0 | ||
DB8RYC* | 9.5 | 2.4 | 9.13 ± 0.25 | 2.50 ± 0.20 | 214.9 ± 3.0 | ||
DB14RC | 9.5 | 2.4 | 9.13 ± 0.25 | 2.50 ± 0.20 | 355.6 ± 3.0 | ||
DB16RC | 9.5 | 2.4 | 9.13 ± 0.25 | 2.50 ± 0.20 | 406.4 ± 3.0 | ||
DB24RC | 9.5 | 2.4 | 9.13 ± 0.25 | 2.50 ± 0.20 | 609.6 ± 3.0 | ||
* Part number consists of a pair of individual tube pieces for double-Y assemblies. Specifications apply to each tube piece.